Skip to content

hBARSCI is now The Lab Stockroom

Glassware Knowledge Center

TC vs. TD Glassware: To Contain vs. To Deliver

TC (to contain), marked "In" under ISO, means glassware holds its stated volume when filled to the mark; TD (to deliver), marked "Ex", means it delivers its stated volume when emptied, so drain time matters. Volumetric flasks are TC, and most pipettes and burettes are TD.

TC vs. TD Key Facts

TC / In To contain: holds the stated volume when filled to the mark
TD / Ex To deliver: dispenses the stated volume when emptied
Marking systems TC and TD are ASTM markings; In and Ex are ISO markings for the same thing
Usually TC Volumetric flasks; many graduated cylinders
Usually TD Volumetric and graduated pipettes, burettes
Why drain time matters TD glassware is calibrated to deliver while draining at a set rate, so rushing or blowing out changes the volume
Reference temperature Volumes are calibrated at a stated temperature, typically 20 °C, and printed on the glass

What Do TC and TD Mean on Glassware?

TC means "to contain": filled to its mark, the vessel holds exactly its stated volume. TD means "to deliver": emptied correctly, it dispenses exactly its stated volume. The difference is the thin film of liquid left on the inside wall after pouring. A TC vessel counts that film as part of its volume; a TD vessel is calibrated so the film stays behind and the right amount still comes out.

New to the terms? Start with What Is Volumetric Glassware?

TC vs. TD and In vs. Ex: Same Meaning, Two Marking Systems

ASTM standards mark glassware TC or TD. ISO standards mark the same calibrations "In" or "Ex." The meaning is identical; only the label differs.

Meaning ASTM marking ISO marking Typical items
To contain TC In Volumetric flasks; many graduated cylinders
To deliver TD Ex Volumetric and graduated pipettes; burettes

Which Glassware Is TC and Which Is TD?

Volumetric flasks are TC: they are used to make up a solution to an exact volume, not to measure a transfer. Most pipettes and burettes are TD, because their job is to move a measured volume into another vessel. Graduated cylinders come both ways, so check the marking before you rely on one for a transfer.

Common TD Mistakes

  • Blowing out a TD pipette. A TD pipette is calibrated to leave a small amount in the tip. Blowing it out adds volume the calibration did not count, unless the pipette is specifically marked for blow-out.
  • Ignoring drain time. TD glassware is calibrated to empty at a set rate. Draining it faster, or pulling it away before it finishes, delivers less than the stated volume.
  • Using a TC flask for a transfer. Pouring from a TC flask leaves the wall film behind, so less than the stated volume comes out.

What the Other Markings on Glassware Mean

Besides TC/In or TD/Ex, volumetric glassware is marked with its volume, its accuracy class (A or B) and its reference temperature, and often its tolerance, the standard it meets and a batch or serial number. The reference temperature is the temperature the volume is calibrated at, typically 20 °C: glass and water both change volume with temperature, so the stated volume is exact only at that temperature. A 1 °C error in water temperature shifts a 100 mL result by about 0.02 mL, one-fifth of the Class A tolerance.

For what the class marking means, see Class A vs. Class B Glassware.

TC vs. TD FAQs

What does "Ex" mean on a pipette?

"Ex" is the ISO marking for "to deliver" (TD under ASTM). The pipette dispenses its stated volume when it drains freely, leaving a small amount in the tip that should not be blown out unless the pipette is marked for blow-out. The Lab Stockroom carries Ex/TD volumetric and graduated pipettes.

What does "In" mean on a volumetric flask?

"In" is the ISO marking for "to contain" (TC under ASTM). Filled to its single mark, the flask holds exactly its stated volume, which makes it the right vessel for making up a solution. The Lab Stockroom stocks In/TC volumetric flasks in Class A and Class B.

Why is 20 °C printed on glassware?

It is the reference temperature: the temperature at which the stated volume is exact. Glass and water both change volume with temperature, so readings taken far from 20 °C drift. The Lab Stockroom's Glassware Knowledge Center explains how this affects calibration.

Can a TC flask be used to deliver a volume?

Not accurately. When a TC flask is emptied, a film of liquid stays on its walls, so less than the stated volume comes out. For a measured transfer, use TD glassware such as a pipette or burette, which The Lab Stockroom stocks in Class A and Class B.

Standards Referenced

Manufacturer Spotlight

Eisco Glassware Markings

Eisco prints the markings this page explains on its volumetric glassware: the volume, the class, TC/In or TD/Ex, and the reference temperature.

Durable MarkingsGraduation ink is tested for tape adhesion, solvent rub, lab washer cycles, autoclaving and thermal shock.
Etched CodeAn etched code on the glass links each piece to its calibration certificate.
Down to 0.025 mLTolerances on Eisco's tightest items.