Volumetric Glassware Glossary
This glossary defines the terms used on volumetric glassware, calibration certificates and lab standards, from Class A and TC/TD to measurement uncertainty, traceability and ISO/IEC 17025. Each definition stands on its own and links to the Glassware Knowledge Center page that covers the term in depth.
Glossary Key Facts
| Terms defined | 24 |
|---|---|
| Sections | Classes and markings; measurement; methods and certificates; standards and accreditation; materials and processing |
| Linking | Each term has its own anchor, so any page can link straight to one definition |
Classes and Markings
- Class A
- The higher accuracy class for volumetric glassware, with about half the tolerance of Class B. Only Class A is individually certified. See Class A vs. Class B Glassware.
- Class B
- The general-purpose accuracy class for volumetric glassware, with a tolerance about twice that of Class A.
- TC / In (to contain)
- A calibration marking meaning the glassware holds its stated volume when filled to the mark. Marked TC under ASTM and In under ISO; volumetric flasks are TC. See TC vs. TD Glassware.
- TD / Ex (to deliver)
- A calibration marking meaning the glassware delivers its stated volume when emptied, so drain time matters. Marked TD under ASTM and Ex under ISO; most pipettes and burettes are TD.
- Reference temperature
- The temperature at which the stated volume is exact, typically 20 °C, printed on the glass. Glass and water both change volume with temperature.
Measurement
- Meniscus
- The curved surface of a liquid in a narrow vessel. For water, read the bottom of the curve at eye level.
- Tolerance
- The amount the true volume is allowed to differ from the stated volume, set by the standard and the accuracy class. See What Is Volumetric Glassware?
- Measurement uncertainty
- A measure of how well a measured value is known: the more widely the values consistent with the measurement are spread, the larger the uncertainty. Tolerance is what the standard allows; uncertainty is how well a calibration knows the actual value.
- Expanded uncertainty
- The combined standard uncertainty multiplied by a coverage factor, k. With k = 2, the stated interval has a level of confidence of about 95 percent, the convention NIST and most calibration laboratories use.
- CMC (calibration and measurement capability)
- The calibration capability a laboratory makes available to customers under normal conditions, as described in its scope of accreditation, including the smallest uncertainty it can offer for a given measurement. The definition is shared by BIPM and ILAC.
- Z factor
- The combined factor used in gravimetric calibration to correct for air buoyancy and convert the mass of water to volume. It depends on water temperature and air pressure and is close to 1.003 mL per gram at typical lab conditions.
- Traceability
- The property of a measurement result whereby the result can be related to a reference through a documented, unbroken chain of calibrations, each contributing to the measurement uncertainty. See the NIST section of Glassware Calibration Certificates.
Methods and Certificates
- Gravimetric calibration
- Calibrating by weighing. The water in or from the glassware is weighed and converted to volume at 20 °C, with corrections for water density, air buoyancy and glass expansion. It is the method in ISO 4787. See How Volumetric Glassware Is Calibrated.
- Graduation
- Placing a volume mark on the glass.
- Calibration
- Measuring the true volume at a mark that already exists. A beaker can be graduated without ever being calibrated.
- Certificate of Conformity
- A document declaring that a batch meets its stated material and specification. Also called a Certificate of Compliance; it does not report measured volumes.
Standards and Accreditation
- ISO 4787
- The ISO standard for testing, calibrating and using volumetric instruments made from glass and plastic, with nominal capacities from 100 µL to 10,000 mL.
- ISO/IEC 17025
- The international standard for testing and calibration laboratories, covering competence, impartiality and consistent operation. See Laboratory Glassware Standards Explained.
- NABL
- India's National Accreditation Board for Testing and Calibration Laboratories. It accredits calibration laboratories to ISO/IEC 17025 and is a signatory to the Global ACI (formerly ILAC) and APAC mutual recognition arrangements. Search NABL accredited laboratories.
- ULR number
- The Unique Laboratory Report number a NABL accredited laboratory puts on each report or certificate it issues within its accredited scope.
- AQL (acceptance quality limit)
- The quality level, stated as percent nonconforming items, that a lot-by-lot sampling plan is indexed to, as in ISO 3951-1. Plans are designed so lots of acceptable quality have a high probability of acceptance.
Materials and Processing
- Borosilicate 3.3
- A low-expansion glass whose properties are defined by ISO 3585. The 3.3 is its coefficient of linear expansion, 3.3 × 10⁻⁶ per °C, which lets it handle direct heat and sudden temperature changes.
- Annealing
- Controlled heating and slow cooling that removes the internal stress left in glass after forming. Borosilicate 3.3 is held near 560 °C, then cooled slowly.
- Grade 3 water
- The general laboratory grade of water under ISO 3696, suitable for laboratory wet chemistry and typically produced by single distillation, deionization, ultrafiltration or reverse osmosis.
Shop Volumetric Glassware
Glossary FAQs
What is the difference between tolerance and measurement uncertainty?
Tolerance is the error the standard allows for the glassware's accuracy class. Measurement uncertainty is how well a calibration knows the actual volume. A calibration certificate reports the measured volume and its uncertainty, along with the tolerance it is judged against. The Lab Stockroom carries Class A glassware with batch or individual certificates.
What does k = 2 mean on a calibration certificate?
It is the coverage factor used to state expanded uncertainty. With k = 2, the stated interval around the measured value has a level of confidence of about 95 percent. The Lab Stockroom's Glassware Knowledge Center explains how calibration certificates report uncertainty.
What is a Z factor in glassware calibration?
It is the factor that converts the mass of water weighed during gravimetric calibration into volume, with a correction for air buoyancy. It changes with water temperature and air pressure and is close to 1.003 mL per gram at typical lab conditions. The Lab Stockroom's Glassware Knowledge Center covers gravimetric calibration under ISO 4787.
What is a ULR number on a NABL certificate?
It is the Unique Laboratory Report number a NABL accredited laboratory gives each report or certificate issued within its accredited scope. The Lab Stockroom carries Eisco glassware, and Eisco issues NABL certificates for select Class A volumetric glassware.
Standards Referenced
- ISO 4787: calibration method for volumetric glassware
- ISO/IEC 17025: competence of testing and calibration laboratories
- ISO 3951-1: sampling plans for inspection by variables, indexed by AQL
- ISO 3585: properties of borosilicate glass 3.3
- ISO 3696: water for analytical laboratory use
Eisco Calibration
Eisco calibrates with Grade 3 water (reverse osmosis plus deionization) in climate-controlled laboratories, and its batch, individual and NABL certificates can be looked up with the code on your glassware.